32
tina88e1 (2014-01)
4.4
Zeroing the Gauge
The gauge is factory calibrated while "standing upright"
(
→
"Calibration Test Report").
We recommend performing a zero adjustment, when the
gauge is operated for the first time.
Due to long time operation or contamination, a zero drift could
occur and zero adjustment may become necessary.
For adjusting the zero, operate the gauge under the same con-
stant ambient conditions and in the same mounting orientation
as normally.
4.4.1
<
ZERO
>
Adjustment
The zero can be adjusted via
•
the <ZERO> button on the gauge,
•
the diagnostic port (
→
[3]),
•
briefly apply the supply voltage to the digital input
"Remote Zero" (pin 10),
•
the RS232 interface (
→
[2], [3]),
•
the Ethernet interface (
→
[1]),
•
the wireless network (
→
[4], [5]).
While the gauge is being heated and/or under atmos-
pheric pressure, the zeroing function is locked in order
for operating errors to be prevented.