12
tina88e1 (2014-01)
Supply
DANGER
The gauge may only be connected to power sup-
plies, instruments or control devices that conform
to the requirements of a grounded protective extra-
low voltage (SELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused.
Supply voltage
at the gauge
Class 2 / LPS
+14 … +30 V (dc) or
±15 V (±5%)
Ripple
≤
1 V
pp
Power consumption
while being heated
at
operating
temperature
≤
20 W
≤
17 W
The gauge is protected against reverse polarity of the supply
voltage and overload.
Electrical connection
15-pin D-Sub, male
Sensor cable
15-pin plus shielding
Cable length
Supply voltage 15 V
≤
4 m (0.14 mm²/conductor)
≤
7 m (0.25 mm²/conductor)
Supply voltage 24 V
≤
20 m (0.14 mm²/conductor)
≤
35 m (0.25 mm²/conductor)
Supply voltage 30 V
≤
40 m (0.14 mm²/conductor)
≤
70 m (0.25 mm²/conductor)
For longer cables, larger conductor cross-sections are required
(R
cable
≤
1.0
Ω
).
Ethernet connection
1×RJ45, 8-pin, socket
Cable
shielded Ethernet cable,
CAT5e quality or higher
Cable length
≤
15 m
Содержание Cube CDGsci
Страница 14: ...14 tina88e1 2014 01 Dimensions mm 110 DN 16 ISO KF DN 16 CF R 8VCR female 110 40 1 4 4 105 190 1 194 5 ...
Страница 15: ...tina88e1 2014 01 15 Damping unit mm 110 36 90 200 188 4 ø5 5 Magnet Weight without damping unit 1 7 kg ...
Страница 20: ...20 tina88e1 2014 01 Carefully remove two covers with e g a small screwdriver Cover Keep the covers ...
Страница 22: ...22 tina88e1 2014 01 and place it on the magnet Magnet ...
Страница 47: ...tina88e1 2014 01 47 Notes ...
Страница 48: ...LI 9496 Balzers Liechtenstein Tel 423 388 3111 Fax 423 388 3700 reachus inficon com t i na88e1 www inficon com ...