28
tina88e1 (2014-01)
4.2
Configuring the network interface
The network interface can be configured via the wired or the
wireless network.
4.2.1 Changing IP Address of the Wired Network
The standard IP address of the wired network (default
192.169.0.248
5)
) can be changed via
•
websites "Configuration" (
→
[5])
•
RS232 ASCII interface with command
IPL [IP address]|[Subnet mask]
(
→
[2]):
e.g. IPL 10.0.0.200|255.255.255.0
•
Ethernet interface with REST services
−
connect Cube with a computer via Ethernet cable (
→
12)
−
set the Ethernet interface of the computer to any IP
address in the subnet 192.168.0.248|255.255.255.0
(e.g. 192.168.0.247)
−
open a web browser (e.g. Mozilla Firefox, version >20, or
Windows Internet Explorer, version
≥
10.0)
−
set the new IP address with the REST services command
http://192.168.0.248/1/cmd/IPL%20
[IP address]|[Subnet mask]
An automatic reset of the gauge is performed, if the IP
address is successfully changed.
4.2.2 Activating the Wireless Network
The wireless network is factory deactivated.
5)
MAC address of the wired network
→
nameplate.
Содержание Cube CDGsci
Страница 14: ...14 tina88e1 2014 01 Dimensions mm 110 DN 16 ISO KF DN 16 CF R 8VCR female 110 40 1 4 4 105 190 1 194 5 ...
Страница 15: ...tina88e1 2014 01 15 Damping unit mm 110 36 90 200 188 4 ø5 5 Magnet Weight without damping unit 1 7 kg ...
Страница 20: ...20 tina88e1 2014 01 Carefully remove two covers with e g a small screwdriver Cover Keep the covers ...
Страница 22: ...22 tina88e1 2014 01 and place it on the magnet Magnet ...
Страница 47: ...tina88e1 2014 01 47 Notes ...
Страница 48: ...LI 9496 Balzers Liechtenstein Tel 423 388 3111 Fax 423 388 3700 reachus inficon com t i na88e1 www inficon com ...