tina88e1 (2014-01)
13
Diagnostic port connection
Jack connector, 3-pin
Cable length
≤
2.5 m
Analog output connection
LEMO plug, 2-pin, with
shielding coaxial
Cable
2-pin plus shielding
Cable length
≤
1 m (0.05 mm²/conductor)
Grounding concept
→
"Power Connection"
Materials exposed to vacuum
ceramics (Al
2
O
3
≥
99.5%),
stainless steel AISI 316L
Internal volume
DN 16 ISO-KF
DN 16 CF-R
8VCR
5.9 cm
3
5.4 cm
3
7.1 cm
3
Admissible pressure (absolute)
100 / 1000 F.S.
1 / 10 F.S.
0.1 F.S.
3 bar
2 bar
1.3 bar
Bursting pressure (absolute)
6 bar
Admissible temperatures
Storage
Operation
–10 °C … +50 °C
+0 °C … +40 °C
Relative humidity
≤
80% at temperatures
≤
+31 °C, decreasing to 50%
at +40°C
Use
indoors only, altitude up to
2000 m NN
Degree of protection
IP 40
Содержание Cube CDGsci