Intended
Use
The temperature compensated Capacitance Diaphragm Gauges
of the CDGsci series are intended for absolute pressure mea-
surement of gases in their respective pressure ranges.
Functional
Principle
A ceramic diaphragm is deflected by pressure. The deflection is
measured capacitively and converted into an analog linear out-
put signal by the digital electronics.
The output signal is independent of the gas type.
Very accurate pressure measurement is achieved by heating the
sensor to a constant temperature of 45°C which results in a
compensation of changes in the ambient conditions and a re-
duced deposition of process products and by-products in pro-
cess applications. An integrated baffle protects the sensor
against coating.
Trademarks
SKY
®
INFICON
GmbH
VCR
®
Swagelok Marketing Co.
Cube™ INFICON
GmbH
Windows CE Microsoft
Patents
EP 1070239 B1, 1040333 B1
US Patents 6528008, 6591687, 7107855, 7140085
Содержание Cube CDGsci
Страница 14: ...14 tina88e1 2014 01 Dimensions mm 110 DN 16 ISO KF DN 16 CF R 8VCR female 110 40 1 4 4 105 190 1 194 5 ...
Страница 15: ...tina88e1 2014 01 15 Damping unit mm 110 36 90 200 188 4 ø5 5 Magnet Weight without damping unit 1 7 kg ...
Страница 20: ...20 tina88e1 2014 01 Carefully remove two covers with e g a small screwdriver Cover Keep the covers ...
Страница 22: ...22 tina88e1 2014 01 and place it on the magnet Magnet ...
Страница 47: ...tina88e1 2014 01 47 Notes ...
Страница 48: ...LI 9496 Balzers Liechtenstein Tel 423 388 3111 Fax 423 388 3700 reachus inficon com t i na88e1 www inficon com ...