6 | Operation
INFICON
78 / 146
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
6.5 Measuring helium ambient concentration
The natural content of helium in the ambient air is about 5.2 ppm.
You can examine the ambient air for an increased helium content (helium
contamination) to determine the appropriate time for further measurements.
ü
The operation mode "Vacuum" is set.
ü
The gas chosen is “helium“
4
ü
There are no long hoses connected to the "purge gas connection". There, the air
enters the device whose helium concentration is measured. See also "Connections
for accessories and control signals [
33]".
ü
The device is in the "Standby" state.
1
Select the cross-hair icon
.
2
Select “I•CHECK“.
3
Press
.
4
View the measurement result.
6.6 Measurement data
You can record measured values in the internal database. Recording can be done
automatically during the measurement or via the "Record measurement data" button
in the measurement screen.
See also "Switch on data recording and configure export [
78]".
You can also export the displayed measured values.
See also "Display measurement cycles [
For the export you can either use a USB flash drive or the network connection, see
also “Export measurement data via network [
80]”.
6.6.1 Switch on data recording and configure export
NOTICE
Reduces the service life of the memory
The recording of many measured values reduces the service life of the internal
memory.
► Disable data recording when you no longer need it.
ü
Operator or Supervisor rights
1
> Setup > Data recording
Содержание 550-520
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Страница 139: ...INFICON Appendix 11 UL6000 Fab Operating instructions iina75en1 04 2109 139 146 11 2 CE Declaration of Conformity ...
Страница 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Страница 143: ...INFICON Index UL6000 Fab Operating instructions iina75en1 04 2109 143 146 Z ZERO 77 Button 26 Factory setting 70 ...
Страница 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Страница 145: ......
Страница 146: ......