6 | Operation
INFICON
62 / 146
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
- TMP
- Filter sniffer tip
- backing pump
- Exhaust filter
- Air filter
2
Adjust.
3
Save
.
6.2.12 Activating or deactivating calibration requests
If the option "calibration request” is not active (factory setting), you only get a
calibration request in the following cases:
• A SL3000 sniffer line with a new serial number has been connected.
• The operation mode has been changed via the operating unit and there was no
calibration in the new operation mode within the last 24 hours.
If the option "calibration request" is active, you will get a calibration request
subsequently in the following cases:
• The period of the device is longer than 30 minutes and the temperature deviation
compared to the last calibration is larger than 5 Kelvin.
• There was no calibration within the last 24 hours in the currently selected
operation mode.
ü
Operator or Supervisor rights
1
> Setup > Measurement > Miscellaneous
2
To switch on further calibration requests activate the option “Calibration
request“. To switch off deactivate this option.
3
Save
.
6.2.13 Extending or reducing operation options
You can always operate the device via touchscreen. Furthermore, you can change the
factory setting, which enables the operation via buttons on the control panel and via
the remote control.
ü
Operator or Supervisor rights
1
> Setup > Measurement > Miscellaneous
2
Under “Local control“ you make your choice in the selection field between:
Содержание 550-520
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Страница 139: ...INFICON Appendix 11 UL6000 Fab Operating instructions iina75en1 04 2109 139 146 11 2 CE Declaration of Conformity ...
Страница 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Страница 143: ...INFICON Index UL6000 Fab Operating instructions iina75en1 04 2109 143 146 Z ZERO 77 Button 26 Factory setting 70 ...
Страница 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Страница 145: ......
Страница 146: ......