Operation | 6
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
61 / 146
ð
If the anticipated pressure conditions are not reached during the evacuation
time, a warning message is shown after the expiration of the evacuation time.
For the pressure conditions, see also the following setting options:
> Setup> Measurement> Pressure limits
and
> Setup > Measurement > Vacuum > Vacuum ranges
6
Save
.
6.2.10 Set maintenance interval "Filter sniffer tip" or "Air filter"
To adjust the maintenance interval to the degree of contamination at the location of
the device, you can choose between a default value or a self-selected maintenance
interval.
ü
Supervisor rights
1
> Setup > Maintenance counter
2
Adjust.
ð
Enter a value of your choice in hours for the "Filter sniffer tip" or press the
adjacent "Default value" button (1000 hours).
ð
Enter a value of your choice in hours for the "Air filter" or press the adjacent
"Default value" button (2500 hours).
3
Save
.
After performing a maintenance, you should reset the desired period until the next
maintenance.
See also
2
Change the filter mat inside the tool box [
2
Replacement of the filter mat on the bottom of the device [
2
Replacing the SL200 sniffer line filter [
2
Replacing the SL3000 sniffer line filter [
6.2.11 Activate or deactivate maintenance requests
Enables or disables warnings when the time for maintenance is exceeded. The time
until the next maintenance is displayed.
ü
Supervisor rights
1
> Setup > Maintenance requests
ð
In the “Enable/disable maintenance requests” window, you can activate or
deactivate the maintenance requests for the following modules:
- Calibration leak
Содержание 550-520
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Страница 139: ...INFICON Appendix 11 UL6000 Fab Operating instructions iina75en1 04 2109 139 146 11 2 CE Declaration of Conformity ...
Страница 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Страница 143: ...INFICON Index UL6000 Fab Operating instructions iina75en1 04 2109 143 146 Z ZERO 77 Button 26 Factory setting 70 ...
Страница 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Страница 145: ......
Страница 146: ......