INFICON
Description | 4
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
19 / 146
4.2.2 Operation mode "Sniffing"
To scan test objects under overpressure with a sniffer probe, you can connect the
SL200 sniffer line or the SL3000 sniffer line.
SL200
The vacuum connection of the sniffer line SL200 is connected to the upper side of the
device on the inlet flange.
The electrical connection of the sniffer line SL200 is connected to the connector
"ACCESSORIES" on the rear of the device, see "Connections for accessories and
33]". An additional cable extension is required for the electrical
connection, see “Accessories and spare parts [
120]“.
SL3000
The sniffer line SL3000 is connected to the connection provided on the rear of the
device, see "Connections for accessories and control signals [
33]". Via this
connector both the electrical connection and the connection of the vacuum system is
made.
As no connection is made to the inlet flange on the upper side of the device, then no
vacuum components that are already connected to the inlet flange need to be
removed when changing to the operation mode "Sniff".
Comparison
SL200
SL3000
Detection limit
< 5 x 10
-8
mbar l/s
< 2 x 10
-7
mbar l/s
Gas flow, distance
sensitivity
25 sccm
160 sccm
Available lengths
3.8 m
3 / 5 / 10 / 15 m
Cable sleeve
Stainless steel
Plastic
Good-Bad-Display
Yes
Yes
ZERO via button
Yes
Yes
Connection on the device
on the inlet flange and via
an electrical connection
via a separate sleeve on
the rear of the device
Vacuum and sniffer line
can be connected at the
same time
-
Yes
Display with measurement
view
-
Yes
Acknowledge faults via
buttons on the sniffer
probe
-
Yes
End standby via a button
on the sniffer probe
-
Yes
Содержание 550-520
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Страница 139: ...INFICON Appendix 11 UL6000 Fab Operating instructions iina75en1 04 2109 139 146 11 2 CE Declaration of Conformity ...
Страница 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Страница 143: ...INFICON Index UL6000 Fab Operating instructions iina75en1 04 2109 143 146 Z ZERO 77 Button 26 Factory setting 70 ...
Страница 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Страница 145: ......
Страница 146: ......