INFICON
Description | 4
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
33 / 146
4.3.4 Connections for accessories and control signals
SNIFFER
Power switch
Connection for
mains cable
Venting connection
Purge gas connection
ACCESSORIES
LD
REMOTE CONTROL
SERVICE
Fig. 6: Connections for accessories and control signals
SNIFFER
For connecting the sniffer line SL3000
ACCESSORIES
Electrical connection of the sniffer line SL200
LD
Connection for bus module BM1000 or I/O module IO1000, maximum*
)
cable
length 30 m (shielded)
REMOTE CONTROL
For connecting the remote control RC1000 via cable, maximum*
)
length 28 m,
or for direct connection of the radio transmitter via an adapter. The remote
control is not part of the scope of delivery of the device.
SERVICE
Connection exclusively for service, maximum*
)
cable length 3 m (shielded)
Connection for mains cable
For power supply of the device
Venting connection
For hoses with an outside diameter of 14 mm, see also “Venting connection
Purge gas connection
For hoses with an outside diameter of 8 mm, see also “Purge gas connection
*
)
In order to avoid the display of incorrect measured values, the maximum cable
lengths specified and the information on shielding must be observed.
NOTICE
The electronics of the device can be destroyed.
► Only connect devices that are provided for the appropriate interfaces.
The connections for the external devices show a safe disconnection from the power
supply and are within the range of the safety extra low voltage (SELV).
Содержание 550-520
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Страница 139: ...INFICON Appendix 11 UL6000 Fab Operating instructions iina75en1 04 2109 139 146 11 2 CE Declaration of Conformity ...
Страница 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Страница 143: ...INFICON Index UL6000 Fab Operating instructions iina75en1 04 2109 143 146 Z ZERO 77 Button 26 Factory setting 70 ...
Страница 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Страница 145: ......
Страница 146: ......