5 | Installation
INFICON
42 / 146
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
CAUTION
Danger of bruising extremities
Feet can be rolled over and crushed.
► Keep feet away from the rollers.
► Do not pull the device.
► Hold the device by the provided handle only and slide it.
► Fix the unit at the destination by locking the wheels with parking brake.
CAUTION
Risk of tripping over the mains cable
► To avoid tripping over an obliquely hanging power cable, untwist the cable
completely or use the “power line strain relief” to guide the mains cable on the
device directly to ground level. See also "Connect the supplied accessories
NOTICE
Operating system can be hacked via USB or Ethernet
The Linux operating system used in the leak detector is not updated automatically and
can therefore contain security vulnerabilities. This vulnerability may be exploited
through the Ethernet and USB interfaces of the leak detector to provide unauthorized
access to the system.
► Ensure that no unauthorized person has access to these interfaces, for example by
using a USB port / Ethernet port lock.
► In order not to jeopardize the security of your company network, never connect the
leak detector directly to the public Internet. This is applies to connections over
WLAN as well as over Ethernet.
► However, if you want to access the web interface of the leak detector remotely, we
recommend an encrypted Virtual Private Network (VPN) connection. However, we
cannot assume any guarantee for the security of VPN connections, which are
provided by third parties.
Prevention of measurement errors due to leaks in the helium source in the
device surroundings
We recommend that you regularly check around the device to a distance of 10 m all
large helium sources for gross leaks. Use a sniffer line to do this.
Содержание 550-520
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Страница 139: ...INFICON Appendix 11 UL6000 Fab Operating instructions iina75en1 04 2109 139 146 11 2 CE Declaration of Conformity ...
Страница 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Страница 143: ...INFICON Index UL6000 Fab Operating instructions iina75en1 04 2109 143 146 Z ZERO 77 Button 26 Factory setting 70 ...
Страница 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Страница 145: ......
Страница 146: ......