INFICON
Installation | 5
UL6000-Fab-Operating-instructions-iina75en1-04-(2109)
47 / 146
Mounting the exhaust hose adapter on the device
• To discharge gases through a DN25 exhaust hose, mount the KF flange DN25
exhaust hose adapter supplied in the exhaust port on the rear of the device. To
seal the exhaust hose adapter, use Teflon tape. Tighten the adapter clockwise.
• If you have an exhaust hose with a hose nozzle (G ½ inch male thread), you can
alternatively fix this hose nozzle in the exhaust port by turning it clockwise.
WARNING
Danger of poisoning due to harmful gases
Depending on the connected container and the gas contained therein, harmful gases
can get into the ambient air via the exhaust filter of the leak detector.
► Make sure you have protection measure to prevent inhaling hazardous gases.
► Do not pump out toxic, corrosive, or explosive gases that create a hazard.
► Connect an exhaust hose. The exhaust gas connection must be continuously
connected to an exhaust gas system and must not be closed.
Содержание 550-520
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 119: ...INFICON Decommissioning 9 UL6000 Fab Operating instructions iina75en1 04 2109 119 146 ...
Страница 139: ...INFICON Appendix 11 UL6000 Fab Operating instructions iina75en1 04 2109 139 146 11 2 CE Declaration of Conformity ...
Страница 140: ...11 Appendix INFICON 140 146 UL6000 Fab Operating instructions iina75en1 04 2109 11 3 RoHS ...
Страница 143: ...INFICON Index UL6000 Fab Operating instructions iina75en1 04 2109 143 146 Z ZERO 77 Button 26 Factory setting 70 ...
Страница 144: ...Index INFICON 144 146 UL6000 Fab Operating instructions iina75en1 04 2109 ...
Страница 145: ......
Страница 146: ......