Product Manual for the Six-Unit Calibration Station for Ventis™
© 2010, 2016 Industrial Scientific Corporation 25
Check V
Cal
Cradle Solenoid X
Or
X-Check Solenoid
Displays to indicate when any solenoid is
not operating properly. The red LED and
the display indicate which cradle numbers
are affected.
Contact factory.
Bump test and calibration
functions are
not
available for
the affected cradles.
Set-up mode is accessible.
USB communications are
accepted and processed.
Check
V
Cal Board
Displays to indicate the board current falls
outside the acceptable limits (or may
indicate an oscillator or A/D failure). All six
red LEDs turn on.
Contact factory.
Bump test and calibration
functions are
not
available.
Set-up mode is accessible.
USB communications are
accepted and processed.
V
Cal Memory
Error
Displays to indicate the station cannot
read from or write to its memory. All six
red LEDs turn on.
Contact factory.
The bump test and calibrate
functions are not available.
Set-up mode is accessible.
USB communications are
accepted and processed.
V
Cal Clock
Error
Indicates an invalid date or clock setting
on the station’s real-time clock. All six red
LEDs turn on.
The user can enter set-up mode
and attempt to re-set the date and
time. (See the manual section, Set-
up Mode.)
Bump test and calibration
functions are not available.
Set-up mode is accessible.
USB communications are
accepted and processed.