© 2010, 2016 Industrial Scientific Corporation 23
The Bump Test Process.
To begin the bump test process, press
BUMP TEST;
hold for three seconds and
release.
NOTE: If a printout of the calibration report is desired, ensure the printer is connected to the station.
DISPLAY
INSTRUCTIONS
X-Warming up
Displays for any instrument that is
charging; indicates the station is
preparing the instrument for bump
testing.
No user action required.
X-Bump in Progress
Displays during the bump test
process. The amber LED turns on for
each affected cradle.
The gas name abbreviation and
calibration gas value for each sensor
display as that sensor is bump tested.
For example, "X-25.0 ppm H2S".
No user action required.
The station reads and applies the bump
test parameters (gas percentage and
response time settings) from the
instrument.
The station determines if any installed
sensor for any instrument is in a
calibration fail or zero fail state. For
those instruments, the station
automatically cancels the bump test
and performs a calibration. If the
calibration is successful, the station
then automatically bump tests the
instrument(s).
X-Bump Passed
X-Bump Failed
Displays along with a red or green
LED to indicate which instruments
have passed (green) or failed (red)
the bump test.
The calibration report for each
instrument is sent out the RS232 port
for printing and is saved to the station.
The station automatically calibrates any
instruments that failed the bump test.
Each cradle's bump test result display
clears when the instrument is removed
from the cradle.