Product Manual for the Six-Unit Calibration Station for Ventis™
© 2010, 2016 Industrial Scientific Corporation 15
Print Events
Instrument X
Displays for each cradle with a
docked instrument; allows the user
to confirm (or cancel) the print
command for each instrument.
Press
BUMP to bypass
printing for the
cradle number shown.
Press
CALIBRATE to send the data
to
the printer for that cradle’s instrument.
Continue to use the BUMP and
CALIBRATE buttons to print or bypass
the printing for each docked
instrument.
If all printing options are bypassed, the
user advances to the next set-up mode
feature,
Print V
Cal Records
.
Printing Instrument
Alarm Events
Displays while the instrument
events print.
No user action required.
NOTE: While the events print, the station ignores all
button presses and USB communications.
Print V
Cal
Records
This procedure sends
all
bump
test and calibration records from
the station memory to the printer.
Press
BUMP to bypass
the procedure;
the user advances to the next set-up mode
feature,
Clear V
Cal Records
.
Press
CALIBRATE to begin
the printing
procedure.
Print Records?
NO YES
Displays to allow the user to
confirm (or cancel) the print
command.
Press
BUMP to cancel
the print
command; the user advances to the next
set-up mode feature,
Clear V
Cal Records
.
Press
CALIBRATE to send the data
to
the printer.
Printing Record
X of YYY
Displays to indicate printing
progress, where X = the record
number currently printing and YYY
= the total number of records that
are being sent to the printer.
Press
BUMP
to can cancel any remaining
printouts.
NOTE: After printing the saved bump test and
calibration reports, they are NOT automatically deleted
from the station. The “Clear V
Cal Records” function is
used to complete that task.