Product Manual for the Six-Unit Calibration Station for Ventis™
© 2010, 2016 Industrial Scientific Corporation 19
Figure 1 lists the information contained in reports that are generated from the
printing procedures described above: printing instrument events and printing
V
Cal Records (calibration and bump test reports).
Report information generated from set-up mode functions.
Print Instrument Events
Print V
Cal Records
Industrial Scientific Corp.
Name of calibration station and its
software version
Date of printout
Instrument serial number
Instrument software version
Instrument hardware version
For each alarm event:
o
Sensor
type
o
Sensor serial number
o
Sensor high alarm threshold
o
Sensor low alarm threshold
o
Peak gas exposure value
during alarm
o
Duration of alarm event in
seconds
o
Time and date the alarm
occurred
o
Instrument user setting
o
Instrument site setting
A blank for the user’s signature
A blank for the user to enter the
date
A blank for the user to enter the
time
Industrial Scientific Corp.
Name of calibration station and its
software version
Date and time of calibration (or
bump test)
Instrument serial number
Instrument software version
Instrument hardware version
Instrument
zero, calibration, or
bump test result (pass or fail)
Recommended date for next
calibration (shown for calibration
only; date is blank for a failed
calibration)
For each sensor:
o
Sensor
type
o
Span reserve (for calibration)
or Final bump test reading (for
bump test)
o
Calibration gas concentration
o
High
alarm
threshold
o
Low alarm threshold
o
Zero, calibration, or bump test
results
A blank for the user to enter the
cylinder lot number
A blank for the user’s signature
Figure 1a.
Event report.
Figure 1b.
Calibration and bump test reports.
###