Product Manual for the Six-Unit Calibration Station for Ventis™
12
© 2010, 2016 Industrial Scientific Corporation
STATION USE
Power-on and -off
To power-on the station, plug the power supply into the power source. To power-
off the station, unplug the power supply. There is no power-on/-off switch.
Start-up Mode
When powered-on, the station performs a series of internal diagnostics as
described below.
If the station fails any diagnostic test, the red LEDs turn on for all affected
cradles. A system error message displays on the LCD to describe each failure
encountered. These messages are defined in the manual section, Status and
Error Messages.
DISPLAY
INSTRUCTIONS
V
·
Cal 6-Unit
Calibration Station
No user action required.
V
·
Cal
v 1.00.07
Displays station name and software version
number (shown: v 3.00.07).
No user action required.
Warming Up
V
·
Cal
Displays during diagnostic testing of the
system’s pumps, solenoids, board, and
memory.
No user action required.
12345678901234567890
ABCDEFGHIJKLMNOPQRST
Displays to verify the correct operation of the
LCD.
No user action required.