72
Level
trOll
Operator’s Manual
0052210 rev. 006
• If available, select the “Wrap data” option (later data will overwrite
earlier data when the memory is full)
• For a log with a scheduled start, select “None” as the stop
condition, or select a stop time that is closer to the start time. You
may intend to stop the log before the scheduled end date arrives,
but the software doesn’t know that.
troubLeShootInG parameter ConFIGuratIon
Problem:
Cannot configure level or other parameters using the Configure
button on the Sensors tab. The Sensor setup screen is shown, but the
Configure... button is dimmed out
Probable Cause 1:
The Level TROLL is actively “polling” (continually
updating real-time readings) in the Home tab
Suggested Remedy:
Return to the Home tab and stop real-time readings
by clicking
Probable Cause 2:
The Level TROLL has an “active” log—a log that
is Ready, Pending, Running, or Suspended as shown in the Status
column of the Logging Tab. Only one “active” log can reside in the
device at a time
Suggested Remedy:
Stop or delete the log if possible. Alternatively,
configure parameters after the log is completed
seCtion 10: trouBLeshootinG