46
RDO 100% Saturation Calibration: Water Saturated Air
Use the procedure below to calibrate the Aqua TROLL 600 RDO sensor, or see the next section for an
alternative method.
a
c
e
b
d
Place the restrictor in calibration
mode (holes near center of
instrument).
Place sponge in restrictor.
Follow the instructions in VuSitu
to finish calibration.
Saturate a small sponge with
water.
Reinstall the end cap and leave
sponge in restrictor for five
minutes.
Sponge
RDO 100% Saturation Calibration: Saturation Bubbler
a
d
e
b
c
f
Fill a 100% saturation bubbler
two-thirds with tap water.
Place sonde into bubbler.
Open the VuSitu mobile app and
tap Calibrations > RDO Saturation.
Follow instructions in VuSitu to
finish calibration.
Turn on bubbler and allow 5-10
minutes for 100% saturation.
Put sonde into deployment mode
by flipping restrictor 180 degrees.
Flip
restrictor
1-970-498-1500
www.in-situ.com