15
Bicarbonate (HCO
3
-
)
440
440
44
Carbonate (CO
3
2-
)
8,600
860
86
Chloride (CI
-
)
7,600
760
76
Dihydrogen Phosphate
(H
2
PO
4
-
)
34,640
3,464
346
Hydrogen Phosphate
(HPO
4
2-
)
34,300
3,430
343
Phosphate (PO
4
3-
)
33,900
3,390
339
Acetate (OAc
-
)
104,200
10,420
1,042
Fluoride (F
-
)
81,400
8,140
814
Sulfate (SO
4
2-
)
685,700
68,570
6,857
Barometric Pressure Sensor Specifications
Conductivity Sensor Specifications
Accuracy
±1.0 mbar
Range
300 to 1,100 mbar
Resolution
0.1 mbar
Sensor Type
Fixed
Response Time
T63<1s, T90<1s, T95<1s
Units of Measure
psi, kPa, bar, mbar, mmHg, inHg
Methodology
Silicon strain gauge
Accuracy
±0.5% of reading plus 1 μS/cm from 0 to 100,000
μS/cm;±1.0% of reading from 100,000 to 200,000
μS/cm; ±2% of reading from 200,000 to 350,000 µS/
cm
Range
0 to 350,000 μS/cm
Resolution
0.1 µS/cm
Sensor Type
Removeable
Response Time
T63<1s, T90<3s, T95<5s
Units of Measure
Actual conductivity: μS/cm, mS/cm Specific
conductivity: μS/cm, mS/cm Salinity: PSU
Total dissolved solids: ppt, ppm Resistivity: Ohms-
cm
Density: g/cm3
Methodology
Std. Methods 2510, EPA 120.1
*Accuracy at calibration points. For greatest accuracy over instrument life, keep the conductivity cell submersed in water for 24-48
hours prior to calibration and deployment.
1-970-498-1500
www.in-situ.com