12
Sensor Specifications
Sensor Summary
Sensors
Shelf Life
Field Life
Recommended
Calibraiton
Frequency
Pressure
Rating - PSI
Usable Depth
Operational
Temperature
Range
pH/ORP
15 months
1 year or greater
10 to 12 weeks
350
200
650
- 5 to 50° C
RDO
NA
2 years or
greater
12 months
350
200
650
- 5 to 50° C
Conductivity
NA
2 years or
greater
User calibration
only if needed
350
200
650
- 5 to 50° C
Temperature
NA
2 years or
greater
NA
350
200
650
- 5 to 50° C
Turbidity
NA
2 years or
greater
User calibration
only if needed
350
200
650
- 5 to 50° C
Pressure
NA
2 years or
greater
User calibration
only if needed
12.8
42.7
108
285
9
30
76
200
30
100
250
650
- 5 to 50° C
Barometric
Pressure
NA
2 years or
greater
User calibration
only if needed
NA
NA
NA
- 5 to 50° C
Ammonium
6 months
6 to 12 months
Monthly
30
25
70
0 to 40° C
Chloride
6 months
1 year or greater
Monthly
350
200
650
0 to 50° C
Nitrate
6 months
6 to 12 months
Monthly
30
25
70
0 to 40° C
Chlorophyll a
NA
2 years or
greater
User calibration
only if needed
350
200
650
- 5 to 50° C
BGA-PC
NA
2 years or
greater
User calibration
only if needed
350
200
650
- 5 to 50° C
BGA-PE
NA
2 years or
greater
User calibration
only if needed
350
200
650
- 5 to 50° C
Rhodamine
NA
2 years or
greater
User calibration
only if needed
350
200
650
- 5 to 50° C
* Shelf life is included in total life of sensor.
** Field life and calibration frequency dependent on site conditions.
1-970-498-1500
www.in-situ.com
Содержание Aqua TROLL 600
Страница 32: ...32 SDI 123 Wire 1 970 498 1500 www in situ com ...
Страница 33: ...33 ModbusMaster 1 970 498 1500 www in situ com ...
Страница 34: ...34 ModbusMasterwithRS232 ConverterRequired 1 970 498 1500 www in situ com ...