Turbidity Sensor Calibration
You must calibrate the turbidity sensor with In-Situ's turbidity standard, or formazine. The app
may not automatically detect the formazine concentration. Instead, a field will appear in which
you can enter the appropriate value. Select
Set User Defined
to begin calibrating with the new
value. If you are using an In-Situ standard and the app does not automatically detect it, perform
the sensor cleaning and maintenance procedure, then select
Retry Auto Detect
.