39
You can’t update instrument firmware while a log is running. Stop a log to update the sonde or
sensors.
Full-Text Messages
The LCD will display text messages instead of status icons when certain conditions are met. Multiple
messages will rotate and display for 3 seconds each.
Text Message
Cause and Remedy
Close battery cover
Battery cover is not fully closed.
Ensure the battery cover is securely tightened.
Install wiper
Wiper sensor port is open.
Install wiper or wiper port plug into center port.
Install sensors
Sensor ports are open.
Install sensors or sensor port plugs.
Install temperature
No Temperature or Conductivity/Temperature sensor detected. Install a
sensor with Temperature.
Install RDO Cap
RDO Cap not detected on RDO sensor. Install RDO cap.
RDO Cap expired
RDO cap is no longer valid.
Install a new RDO Cap.
RDO Cap XXX days
Temporary text message update on the lifespan of the RDO Cap.
Starting a Log with the LCD Menu
Stopping a Log
The data log menu allows you to create and start a basic linear data log if one doesn’t already
exist on the instrument.
Activate the LCD menu and select
Data Log
.
Set the logging interval.
Choose
New
.
The log starts immediately.
a
c
b
d
1-970-498-1500
www.in-situ.com