42
Micro SD Card
The Aqua TROLL 600 uses a micro SD card for data storage and updating sonde firmware. You can remove
the SD card and replace it with another for data download, or use the same card. An SD card is not required
to log data.
Removing the Micro SD Card
Updating Firmware
Downloading and Deleting Data from the Micro SD Card
The micro SD card is housed in a slot near the positive battery terminal and the twist-lock connector.
1 Remove the battery cover from the end of the instrument.
2 Push the micro SD card in towards the instrument to release the card.
3 Remove the card from the slot.
4 To reinstall the card, note the card position diagram in the finger slot. Push the card into the slot until you
hear a click.
1 Remove card from the sonde and insert it into a micro SD card adapter.
2 Insert the adapter into a PC or laptop.
3 Open the micro SD card using the file explorer.
4 Open the folder titled “Serial Number.LOG.” For example, “424690.LOG.”
Logs are named by date and then number. For example, a log started on November 12, 2015
would be named “15111200.CSV” - 15 is the year, 11 is the month, 12 is the day, and 00 is the
log number. If multiple logs are recorded on the same day the last number will increase by one
for each consecutive log.
5 Select the logs you wish to download and move them to the destination.
6 Select the logs you wish to delete and press Delete on the keyboard. Deleted logs cannot be recovered.
1 Download the firmware update files to your computer.
2 Remove the SD card from the unit and insert it into your computer.
You may need a micro SD card adapter to connect the SD card to your computer. One is
included with your Aqua TROLL 600, but any micro SD card adapter can be used.
3 Copy the firmware update files into the ISI.FW file on the SD card.
4 Install the SD card into the instrument and install the battery cover.
5 Turn on the LCD screen by inverting the sonde so the sensor end faces up.
6 Access the instrument menu by tapping the Aqua TROLL 600 logo on the yellow label three times. For
more information on accessing menus, see “LCD Screen.”
7 Scroll to Updates by tilting the instrument. Access the Updates menu by tapping the Aqua TROLL 600
1-970-498-1500
www.in-situ.com