45
To start the measurement of the material press the
button „Measurement“
. To improve the accuracy,
four measurments will be taken. The individual values
are averaged. The duration of a measurement takes
around 20 seconds. After each measurement the signal
runtime will be displayed for a short period.
NOTE:
Ensure that the probe rods are totaly covered with the material during the time of the measurement and that
the probe is not moved.
Attention:
If you choose “SAVE” at the end of calibration, the stored calibration in the probe electronics will be overwritten!
The original calibrations can be restored by connecting the probe with a RS485 adapter (for example SM-USB)
to a PC and to use the software PICO-Config.
Finally, you can store the calibration into the choosen
calibration storage inside the probe. Choose „SAVE“ and
confirm by pressing the button „Measurement“
.
If you choose “DISCARD” everything is left untouched.
To name of the changed calibration, is the original with
the prefix „OWN:“.
TIPP:
Save your calibrations which are stored inside the probe to your PC. A RS485 adapter (such as SM-USB) and the
software PICO-Config is required. With this method, individual calibrations can also be transferred to other probes.