D
x
Telemetry System
Manual Vers. 2.7
88
/
92
15. Switch on SCT 3 and proceed with the transmitter connection as described in 3-8.
However, select a different transmission frequency (e.g.,
freq2
= 870 MHz).
16. Switch on SCT4 and proceed as described in section 10. Select the same
transmission frequency as for SCT3 (e.g.,
freq2
= 870 MHz).
17. Change the reception frequency of the RCI2 as described under 11. to the new
transmission frequency
freq2
18. Test whether SCT3 and SCT4 are found as described in section 13.
19. The system is ready for operation.
Caution:
SCT1 and SCT2 are now configured via RCI1, SCT3 and SCT4 are
configured via RCI2. If you want to change channel parameters such as the sampling
rate, both systems must be configured separately.
10.4 Unstable signal, when connecting more than one SCT
It’s often necessary, after programming the frequency and “logical number”, to program a
channel on each SCT in order to receive interference-free reception.
Содержание CAEMAX DX
Страница 31: ...Dx Telemetry System Manual Vers 2 7 31 92 Figure 20 Strain gauge channel configuration...
Страница 44: ...Dx Telemetry System Manual Vers 2 7 44 92 Menu tree continued...
Страница 84: ...Dx Telemetry System Manual Vers 2 7 84 92 10 Questions Answers 10 1 Inductive Power Supply Troubleshooting...
Страница 85: ...Dx Telemetry System Manual Vers 2 7 85 92...
Страница 86: ...Dx Telemetry System Manual Vers 2 7 86 92...