10.2 Integration of SCTs
If problems with your telemetry systems occur, it can be necessary to reintegrate the SCTs.
In that case, please regard the following instructions.
1. Turn the D
X
RCI on.
2. With
Devices → Device X
, delete all registered devices with the
Delete
button.
3. Disconnect all SCTs from the power supply, except the one to be integrated.
4. Integrate the active SCT (see section 3.5).
5. Repeat step 4 until all SCTs are integrated.
10.3 Changing the transmission frequency of SCT and RCI
If two DX receiving units (RCIs) are operated in parallel, different transmission frequencies
(here referred to as freq1 and freq2) must be set for each receiving unit and corresponding
transmitting units (SCTs).
The following example describes the configuration of a system with two RCIs, RCI 1 and RCI
2, as well as 4 SCTs. SCT1 and SCT2 are operated with RCI1, SCT3 and SCT4 with RCI2
1. Switch the receiving unit 1
on
(RCI 1) and the receiving unit 2
off
(RCI 2).
2. Turn off
all
transmission units (SCTs) except for the SCT you wish to connect (i.e.,
interrupt the power supply).
3. Set the RCI to the same frequency as the SCT to which you want to establish a
connection (in the menu under
Devices → Base RCI → Radio Frequency
). Default
setting at delivery: 864 MHz.
4. Locate SCT 1 (
Devices → Device 1 → Search
). After the message
Device found!
Appears. Continue with the configuration.
5. Assign the logical number 1 (under
Devices → Device 1 → Logical Number
, see
section 3.5)
6. Change the transmission frequency of SCT 1 (under
Devices → Device 1 →
Frequency
) to the desired transmission frequency
freq1
. For a trouble-free operation,
the transmission frequencies should be as far apart as possible (i.e.,
freq1
= 864
MHz for RCI 1 and
freq2
= 870.0 MHz for RCI 2)
7. Program the SCT 1 (
Devices → Device 1 → Program
).
Attention
: Due to the changed transmission frequency, the SCT is no longer
available!
8. After the message
Program done!
Appears. Turn off the SCT1 (disconnection from
the power supply).
9. Now turn on the SCT2.
10. Repeat steps 3-8 with the SCT 2. Similar to the step described in 5, now define the
logical number 3.
11. Change the transmission frequency of the RCI (
Devices → Base RCI
) to the
frequency set at SCT1 and SCT2
freq1
. Confirm with
Program
.
12. Turn both SCTs on.
13. Test whether SCT 1 and SCT 2 are found with
Devices → RCI X → Search
. Leave
the menu with OK.
14. Switch off RCI 1, SCT1 and SCT2 again and start RCI 2.
Содержание CAEMAX DX
Страница 31: ...Dx Telemetry System Manual Vers 2 7 31 92 Figure 20 Strain gauge channel configuration...
Страница 44: ...Dx Telemetry System Manual Vers 2 7 44 92 Menu tree continued...
Страница 84: ...Dx Telemetry System Manual Vers 2 7 84 92 10 Questions Answers 10 1 Inductive Power Supply Troubleshooting...
Страница 85: ...Dx Telemetry System Manual Vers 2 7 85 92...
Страница 86: ...Dx Telemetry System Manual Vers 2 7 86 92...