D
x
Telemetry System
Manual Vers. 2.7
46
/
92
4.2.2 Menu: Device
Device1
Serial number:
This number is used for addressing the SCT. Only this device is
addressed when programming.
Search:
This function is used to search for the respective frequency according
to the SCT. The “serial number” and the "logical number" are
determined and entered into the field.
IMPORTANT:
When searching, make sure that only one SCT is
turned on to provide a unique identification
.
Version:
Information about the SCT version.
Please note:
Current firmware version is displayed after pressing
Search.
Radio frequency:
You can set the desired radio frequency through a drop-down menu.
The SCTs can only communicate with RCIs of the same frequency.
Up to 4 SCTs of the same frequency can be operated via one RCI.
Each carrier frequency offers a bandwidth of approx. 6 kS/s. The
modification of this setting is only activated after programming and
rebooting the SCT.
IMPORTANT: Always set the SCTs to the desired frequency first.
Otherwise you will “lock yourself out”.
Error correction:
An additional hardware error correction can be activated when using
the 2.4 GHz version. This feature will reduce the signal bandwidth.
Packet mode:
Reserved for future settings.
Logical number:
This number defines the starting point for the data transmission of the
respective SCT in the corresponding timeslot. This time domain is
subdivided into 4 secti
ons (“logical number” 1-4).
If you only use one SCT, set the “logical number” to “1”. By doing so,
you are able to use the whole timeslot (1 - 4).
If you operate with two SCTs use numbers 1 and 3; by doing so,
each SCT has 2/4 of the timeslots available.
When using 3 or 4 SCTs the numbers can be allocated as you like.
IMPORTANT: At any given frequency, the "Logical Number" can only
be assigned once.
Содержание CAEMAX DX
Страница 31: ...Dx Telemetry System Manual Vers 2 7 31 92 Figure 20 Strain gauge channel configuration...
Страница 44: ...Dx Telemetry System Manual Vers 2 7 44 92 Menu tree continued...
Страница 84: ...Dx Telemetry System Manual Vers 2 7 84 92 10 Questions Answers 10 1 Inductive Power Supply Troubleshooting...
Страница 85: ...Dx Telemetry System Manual Vers 2 7 85 92...
Страница 86: ...Dx Telemetry System Manual Vers 2 7 86 92...