Load the Flow Cell
1
Open the flow cell compartment door.
2
Press the release button to the right of the flow cell latch.
Figure 17
Open Flow Cell Latch
3
If present, remove the used flow cell from a previous run.
4
Make sure that the flow cell stage is clean. If debris is present, clean the flow cell stage with an alcohol
wipe.
5
Place the flow cell on the flow cell stage over the alignment pins.
Figure 18
Place Flow Cell on Stage
6
Close the flow cell latch to secure the flow cell.
Figure 19
Close Flow Cell Latch
7
Close the flow cell compartment door.
Load the Reagent Cartridge
1
Open the reagent compartment door.