6
[Optional]
If you are using custom primers, select the checkbox for the primers used.
u
Read 1—Custom primer for Read 1.
u
Index 1—Custom primer for Index 1.
u
Index 2—Custom primer for Index 2.
u
Read 2—Custom primer for Read 2.
7
[Optional]
Select Advanced Settings to change settings for the current run.
u
Purge consumables for this run—Change the setting to purge consumables automatically after the
current run.
u
Use run monitoring for this run—Change the setting to use run monitoring in BaseSpace. When
enabled, a BaseSpace login and internet connection are required.
u
Recipe—Select a custom recipe for this run.
u
Output folder—Change the output folder location for the current run. Select Browse to navigate to a
folder location.
8
Select Next.
NOTE
Do not open the reagent compartment door or the flow cell compartment door during the automated
check or during the sequencing run.
Review Automated Check
1
Review the automated check results.
u
To stop a check in progress, select Cancel.
u
For any items that do not pass, an action is required before you can proceed. See
.
u
To restart the check, select Retry. The check resumes at the first incomplete or failed check.
2
To start the run, select from the following options.
u
If the system is not configured to start automatically after a successful check, select Start.
u
If the system is configured to start automatically after a successful check, the sequencing run begins
automatically. You do not have to be present. However, if any errors occur during the check, the run
does not begin automatically.
Set Up a Run (BaseSpace or Local Run Manager Configuration)
1
From the Home screen, select Sequence.
The Sequence command releases consumables from a previous run and opens the series of run setup
screens.
Log in to BaseSpace or Local Run Manager
1
Enter your user name and password.
2
Select Next.
Select Available Run
1
Select a run name from the list of available runs.
Use the up and down arrows to scroll through the list or enter a run name in the Search field.
2
Select Next.
Document # 1000000002695 v02 Material # 20014309
For Research Use Only. Not for use in diagnostic procedures.
24
MiniSeq System Guide