125
returned to the calibration mode screen.
9.12
Calibration of Peak Current
Use the
,
soft keys to navigate to the A-Peak xx.xA parameter and press the
Select soft key.
Follow the prompt message provided on the display, and press the Test/Reset
button to move into the calibration screen for current. If you press the Exit soft
key at this screen you return to the calibration mode screen.
Enter the current reading from the ammeter with the numeric keypad. When the
value has been selected press the Enter soft key and you will be moved to the
next calibration parameter Volt 150.0V. The soft keys Volt+ and Volt- are
available, if needed, to adjust the voltage output of the instrument to set a proper
peak current value if a non-recommended load is used. If you press the Esc soft
key you will be returned to the calibration mode screen.