103
from the instrument. This command set also includes query commands. These
query commands will retrieve data from the instrument. The GPIB bus application
requires an IEEE-488 read command to be sent after the query command. These
commands include functions for retrieving test data, test results and metering
values.
Command
Description
AC Program
Mode
AC Manual
Mode
DC Program
Mode
DC Manual
Mode
Unit
TEST
TEST
Power On
RESET
RESET
Power Off
TD?
Testing meters
data
Memory, Step, Status,
Freq, Volt, Curr,
Power, APeak, PF, Q,
CF, VA, Timer
Memory, Status,
Freq, Volt, Curr,
Power, APeak, PF,
Q, CF,VA
Memory, Step, Status,
Volt, Curr, Power,
Timer
Memory, Status,
Volt, Curr, Power
RD XX?
Results meters
data
TDFREQ?
Testing
frequency meter
40-1000
Hz
TDVOLT?
Testing voltage
meter
0.0~300.0
0.0~420.0
V
TDCURR?
Testing current
meter
See Note1-1
See Note1-1
See Note2-1
See Note2-1
A
TDAP?
Testing current
peak meter
See Note4
A
TDP?
Testing power
meter
See Note3-1
W
TDPF?
Testing pf meter
0.000~1.000
TDQ?
Testing Q meter
See Note3-1
TDCF?
Testing CF
meter
0.00~10.00
TDVA?
Testing VA
meter
See Note3-1
VAR
TDTIMER?
Testing timer
meter
0.0~999.9
0.0~999.9
s/m/h
METER X
SELECT
METER X
X=0-8,0=F,1=A,2=P,
3=Q,4=CF,5=AP,6=V
A,7=PF,8=CYCLE
X=0-
7,0=F,1=A,2=P,
3=Q,4=CF,5=AP,6
=VA,7=PF
X=1,2,8
1=A,2=P,8=CYCLE
X=1,2
1=A,2=P
METER?
SELECT
METER?
SDTRG
TRANSIENT
TRIGGER
Trigger one time
TRANSIENT
STEPCYCLE?
STEP CYCLE?
0~9999
0~9999
MEMORYCYCLE?
MEMORY
CYCLE?
0~9999
0~9999
LOOPCYCLE?
LOOP CYCLE?
0~9999
0~9999
RR?
READ:RESET
See Note 5
RI?
READ:INTERL
OCK
0~1, 0=CLOSE,
1=OPEN
EXTP
EXT TRIGGER
pulse
Trigger one pulse on
SYNC
TEST
Turns on the output voltage at the selected step loaded into memory.
RESET
Turns the output voltage off or resets the instrument in the event of a failure.