The decimal separator in a number is a dot (code: hex 0x2E).

The above details correspond as far as possible to the recom-
mendations of the NAMUR working party (NAMUR recom-
mendations for the design of electrical plug connections for 
analogue and digital signal transmission on individual items of 
laboratory control equipment, rev. 1.1).

Commands

Function

IN_PV_2

Read the internal actual temperature

IN_PV_1

Read the external actual temperature

IN_SP_2

Read the internal setting temperature

IN_SP_1

Read the external target temperature

IN_SP_3

Read the safety setting temperature

IN_SP_4

Read the fan speed value

IN_SP_40

Read the flap open value 

OUT_SP_2  X

Set the internal setting temperature, X = 0 to 
250 in steps of 0.1 °C

OUT_SP_12@n Setting WD2 safety limit temperature with set 

value echo

OUT_SP_4  X

Fan target value, X = 0 to 100 in steps of 10 %

OUT_SP_40 X

Flap target value, X = 0 to 100 in steps of 10 %

OUT_WD1@m Watchdog mode 1: if event WD1 should occur, 

the heating functions are switched off and WD1 
is displayed. Set watchdog time to m (20 - 1,500) 
seconds, with watchdog time echo. This com-
mand launches the watchdog function and must 
be transmitted within the set watchdog time.

The NAMUR commands and the additional specific 

IKA

®

 com-

mands serve only as low level commands for communication be-
tween the device and the PC. 
With a suitable terminal or communications programme these com-
mands can be transmitted directly to the device. The 

IKA

®

 software 

package, Labworld

soft

®

, provides a convenient tool for controlling 

the device and collecting data under MS Windows, and includes 
graphical entry features, for motor speed ramps for example.

OUT_WD2@m Watchdog mode 2: if event WD2 should occur, 

the temperature target value is changed to the 
WD safety temperature limit value. The warning 
WD2 is displayed. The WD2 event can be reset 
with OUT_WD2@0 - this also stops the watchdog 
function. Set watchdog time to m (20 - 1,500) 
seconds, with watchdog time echo. This com-
mand launches the watchdog function and must 
be transmitted within the set watchdog time.

START_2

Start the heating function

STOP_2

Stop the heating function

RESET

Switches off the device function

 NOTICE

When the USB cable is disconnected 

while the device is running, the 

device stays in PC mode. To reset 

to function without PC, switch the 

device off and on again.

USB B

USB A

1

4
3

2

1
2
3
4

“Watchdog” function, monitoring the serial data flow:
The following applies to situations where the watchdog function is 
enabled (see Namur instructions). If no new transmissions of these 
commands from the PC take place within the preset watchdog 
time, the heating function will be shut down according to the 
watchdog mode selected or will be controlled using the preset 
setpoints. An operating system crash, a PC power failure or a fault 
in the connecting cable to the device can cause an interruption in 
data transmission.

“Watchdog”– Mode 1
If an interruption in data transmission occurs which is longer than 
the preset watchdog time, the heating function will be shut down 
and the error message PC 1 will be displayed.

“Watchdog”– Mode 2
If an interruption in data transmission occurs which is longer than 
the preset watchdog time, the temperature setpoint will be set to 
the WD safety temperature setpoint. The PC 2 warning message 
will be displayed.

Connections between device and external devices:

USB 2.0 cable A - B:
This cable is required to connect USB interface to a PC

Содержание Oven 125 control - dry glass

Страница 1: ...20000023854 Operating instructions EN DO 125 control_032019 IKA Oven 125 control dry IKA Oven 125 control dry glass...

Страница 2: ...nterface RS323 7 Mains socket 8 Venting flap 9 Cable feed through external temperature sensor 10 Oven light 11 Fan 12 Rack slide in supports Device setup Dangerous spots Dangerous spots Device setup F...

Страница 3: ...rature monitoring 7 Transportation and storage 7 Unpacking 8 Installation 8 Operator panel and display 11 Setting the temperature limit 12 Commissioning 12 Menu settings 13 Menu structure 13 Navigatio...

Страница 4: ...ospheres it is not EX protected With substances capable of forming an explosive mixture appropriate safety measures must be applied e g working under a fume hood To avoid bodily injury and property da...

Страница 5: ...will automatically restart in mode B and C follow ing any interruption to the power supply Protect the power cable against kinks Check power cables at regular intervals for possible material fatigue M...

Страница 6: ...is operated with accessories that are not supplied or recommended by IKA If the device is operated improperly or in contravention of the IKA specifications If the device or the printed circuit board...

Страница 7: ...eeds hysteresis limit and recovers If the user did not select the option to turn OFF heating operation whenever the temperature overshoots or undershoots the stable set temperature and passes the pred...

Страница 8: ...ble Setting up Fixing against the wall is compul sory Place the device on and even clean dry and fireproof surface in well ventilated environment Fixing the device to the wall NOTICE The screws used f...

Страница 9: ...ad shelves evenly Stacking the device CAUTION Never stack more than two devices When stacking the device four per sons are required to lift it B NOTICE In order to keep enough space please use stackin...

Страница 10: ...weight of the door about 15 kg WARNING Open the door Hold the door and loosen the hinge pin with a suitable tool Remove the door carefully Alternate the position of the keepers and hinges with a 4 mm...

Страница 11: ...or ventilation flaps open K Graph selection button Graph function display Symbols at working screen Symbol Description Key lock is activated USB storage medium is connected USB B is connected Light is...

Страница 12: ...ing c Fan target speed d Counter Timer display e Temperature target value f Actual oven temperature g Actual temperature of inserted external temperature sensor h Timer target value a b c d e f g h If...

Страница 13: ...n Timer Counter Timer Days Hours Minutes Controller Time Temperature activated Operating mode A B C activated Program Program 1 5 Start Loop count Endless loop Immediately Delayed Immediately Delayed...

Страница 14: ...3 point calibration Actual temperature note Gives a notification when the actual temperature is in the range of 1 0 K of the target temperature and the symbol is shown next to the target temperature T...

Страница 15: ...matrix of lines segments and columns functions Seg No column Segment Number Temp column specifies the desired target temperature Control Mode column The temperature time profiles can be created using...

Страница 16: ...ew of the device temperature parameters Software Displays the device software version used Operation Temperature Temperature control NOTICE The temperature calibration and comparison for the external...

Страница 17: ...n under the fan speed 3 Press and turn the rotary push knob E to set the speed in 10 steps 0 stopped 100 full speed 4 Press the rotary push knob E The value is saved as a mea surement 5 Press the back...

Страница 18: ...recognised Use the USB interface in conjunction with labworldsoft for opera tion in Remote mode and also to update the firmware USB device drivers First download the latest driver for IKA devices wit...

Страница 19: ...for motor speed ramps for example OUT_WD2 m Watchdog mode 2 if event WD2 should occur the temperature target value is changed to the WD safety temperature limit value The warning WD2 is displayed The...

Страница 20: ...tion method than the method defined by IKA is planned you must ascertain from IKA that this method does not damage the device Spare parts order When ordering spare parts please specify Machine type Se...

Страница 21: ...external temperature sensor correctly Contact the IKA Service Department Temperature in the device too high PCB power a Ambient temperature PCB too high b PCB malfunction Heating off Turn off the dev...

Страница 22: ...control dry kg 89 Oven 125 control dry glass 2 racks Permitted ambient temperature range C 5 40 Permissible relative humidity 80 Protection class according to DIN EN 60529 IP 20 RS 232 interface yes U...

Отзывы: