background image

12

Setting the safety temperature limit

The maximum temperature that can be reached in the oven is lim-
ited by an adjustable safety temperature limiter. Once this limit has 
been attained, the device switches off the heating. 
Adjust the temperature limit (increase the value by turning clock-
wise / decrease the value by turning anticlockwise) using the 
screwdriver supplied with the device.

The set safety temperature is displayed in the operating field (F).

Temperature limit
Factory setting: approximate max. value
Adjustment range: see “Technical data”.

WARNING

The safe temperature limit must 

always be set to at least 25 °C

lower than the flash point of the 

media used.

Commissioning

F

Check whether the voltage specified on the type plate matches the 
mains voltage available.

The power socket used must be earthed.

Pos.  Description
a       

Time

b       

Ventilation flap opening

c       

Fan target speed

d       

Counter/Timer display

e       

Temperature target value

f        

Actual oven temperature

g       

Actual temperature of inserted 

           external temperature sensor

h       

Timer target value

a

b

c

d

e

f

g

h

If the above conditions are met, the device is ready for operation 
after plugging in the mains plug. If these conditions are not met, 
safe operation is not guaranteed and the device could be damaged.
Observe the ambient conditions (temperature, humidity, etc.) listed 
under “Technical data”.

Switch on

After switching on (Fig. 1, 1) the start screen displays the firmware 
version and device type, device designation, and user-defined de-
vice name.

Firmware Update Tool
After the start screen, the information start screen for the Firmware 
Update Tool appears. You can use the tool to update the software of 
your IKA device to the latest version. Firmware updates contain 
new functions or optimisations of previous functions. 

Working screen

After displaying the start screen and the firmware update 
information, the working screen appears. The device is ready for 
operation.

+

-

Содержание Oven 125 control - dry glass

Страница 1: ...20000023854 Operating instructions EN DO 125 control_032019 IKA Oven 125 control dry IKA Oven 125 control dry glass...

Страница 2: ...nterface RS323 7 Mains socket 8 Venting flap 9 Cable feed through external temperature sensor 10 Oven light 11 Fan 12 Rack slide in supports Device setup Dangerous spots Dangerous spots Device setup F...

Страница 3: ...rature monitoring 7 Transportation and storage 7 Unpacking 8 Installation 8 Operator panel and display 11 Setting the temperature limit 12 Commissioning 12 Menu settings 13 Menu structure 13 Navigatio...

Страница 4: ...ospheres it is not EX protected With substances capable of forming an explosive mixture appropriate safety measures must be applied e g working under a fume hood To avoid bodily injury and property da...

Страница 5: ...will automatically restart in mode B and C follow ing any interruption to the power supply Protect the power cable against kinks Check power cables at regular intervals for possible material fatigue M...

Страница 6: ...is operated with accessories that are not supplied or recommended by IKA If the device is operated improperly or in contravention of the IKA specifications If the device or the printed circuit board...

Страница 7: ...eeds hysteresis limit and recovers If the user did not select the option to turn OFF heating operation whenever the temperature overshoots or undershoots the stable set temperature and passes the pred...

Страница 8: ...ble Setting up Fixing against the wall is compul sory Place the device on and even clean dry and fireproof surface in well ventilated environment Fixing the device to the wall NOTICE The screws used f...

Страница 9: ...ad shelves evenly Stacking the device CAUTION Never stack more than two devices When stacking the device four per sons are required to lift it B NOTICE In order to keep enough space please use stackin...

Страница 10: ...weight of the door about 15 kg WARNING Open the door Hold the door and loosen the hinge pin with a suitable tool Remove the door carefully Alternate the position of the keepers and hinges with a 4 mm...

Страница 11: ...or ventilation flaps open K Graph selection button Graph function display Symbols at working screen Symbol Description Key lock is activated USB storage medium is connected USB B is connected Light is...

Страница 12: ...ing c Fan target speed d Counter Timer display e Temperature target value f Actual oven temperature g Actual temperature of inserted external temperature sensor h Timer target value a b c d e f g h If...

Страница 13: ...n Timer Counter Timer Days Hours Minutes Controller Time Temperature activated Operating mode A B C activated Program Program 1 5 Start Loop count Endless loop Immediately Delayed Immediately Delayed...

Страница 14: ...3 point calibration Actual temperature note Gives a notification when the actual temperature is in the range of 1 0 K of the target temperature and the symbol is shown next to the target temperature T...

Страница 15: ...matrix of lines segments and columns functions Seg No column Segment Number Temp column specifies the desired target temperature Control Mode column The temperature time profiles can be created using...

Страница 16: ...ew of the device temperature parameters Software Displays the device software version used Operation Temperature Temperature control NOTICE The temperature calibration and comparison for the external...

Страница 17: ...n under the fan speed 3 Press and turn the rotary push knob E to set the speed in 10 steps 0 stopped 100 full speed 4 Press the rotary push knob E The value is saved as a mea surement 5 Press the back...

Страница 18: ...recognised Use the USB interface in conjunction with labworldsoft for opera tion in Remote mode and also to update the firmware USB device drivers First download the latest driver for IKA devices wit...

Страница 19: ...for motor speed ramps for example OUT_WD2 m Watchdog mode 2 if event WD2 should occur the temperature target value is changed to the WD safety temperature limit value The warning WD2 is displayed The...

Страница 20: ...tion method than the method defined by IKA is planned you must ascertain from IKA that this method does not damage the device Spare parts order When ordering spare parts please specify Machine type Se...

Страница 21: ...external temperature sensor correctly Contact the IKA Service Department Temperature in the device too high PCB power a Ambient temperature PCB too high b PCB malfunction Heating off Turn off the dev...

Страница 22: ...control dry kg 89 Oven 125 control dry glass 2 racks Permitted ambient temperature range C 5 40 Permissible relative humidity 80 Protection class according to DIN EN 60529 IP 20 RS 232 interface yes U...

Отзывы: