General information:

• 

Read the operating instructions completely before 

starting up and follow the safety instructions.

• Keep the operating instructions in a place where it can be 

accessed by everyone.

• Ensure that only trained staff work with the device.
• Follow the safety instructions, guidelines, occupational 

health, safety and accident prevention regulations.

• Set up the device in a spacious area on an even, stable, 

clean, non-slip, dry and fireproof surface.

• Do not operate the device in explosive atmospheres, 

with hazardous substances.

• Protect the device and accessories from bumping and 

impacting. 

• Check the device and accessories for damage before 

each use. Do not use damaged components.

• Safe operation is only guaranteed with the accessories 

described in the ”

Accessories

” section.

• The socket for the mains cord must be easily accessible.
• Socket must be earthed (protective ground contact).
• The voltage stated on the type plate must correspond to 

the mains voltage.

•  The device can only be disconnected from the mains sup-

ply by pulling out the mains plug or the connector plug.

• Disconnect the mains plug before attaching or changing 

any accessories.

• Disconnect the mains plug before cleaning, maintenance 

and transportation of the unit.

• The device must only be opened by trained specialists, 

even during repair. The device must be unplugged from 
the power supply before opening. Live parts inside the 
device may still be live for some time after unplugging 
from the power supply.

Coverings or parts that can be 
removed from the device with-
out the aid of any tools must 

be put back on the device again to ensure safe opera-
tion, for example to keep foreign objects and liquids, 
etc. from getting into the device.

• The device may only be used as prescribed and as de-

scribed in the operating instructions. This includes op-
eration by instructed specialist personnel.

• When using critical or hazardous materials in your pro-

cesses, 

IKA

®

 recommends to use additional appropriate 

measures to ensure safety in the experiment. For exam-
ple, users can implement measures that inhibit fire or 
explosions or comprehensive monitoring equipment. 

• Process pathogenic material only in closed vessels under 

a suitable fume hood. Please contact 

IKA

®

 application 

support if you have any question.

If the 

OFF

 switch is not within 

reach when device is operating, 
an 

EMERGENCY STOP

 switch 

that can be easily accessed must 
be installed in the work area.

• A laboratory circulator heats and circulates fluid accord-

ing to specified parameters. This involves hazards due to 
high temperatures, fire and general hazards due to the 
device of electrical energy. The user is largely protect-
ed by the device of relevant standards. Further hazard 
sources may arise due to the type of tempering fluid, 
e.g. by exceeding or undercutting certain temperature 
thresholds or by the breakage of the container and re-
action with the heat carrier fluid. It  is  not  possible  to 
consider all eventualities. They remain largely subject to 
the judgment and responsibility of the operator. For this 
reason, it may become necessary for user to take other 
precautionary safety measures. 

•  Insufficient ventilation may result in the formation of explo-

sive mixtures. Only use the device in well ventilated areas.

The safety circuit (safe tem-
perature) must be adjusted so 
that the maximum permissible 

temperature cannot be exceeded even in the event of a 
fault. Check the safe temperature circuit on a regular ba-
sis (see section “

Setting the safety temperature

“) . 

• Securely fix the

 

ICC 

immersion circulator for use at the 

bath, so that it cannot tip over.

• When device is used for external circulation, extra pre-

caution must be taken for hot fluid leakage due to dam-
aged hose.

• Use suitable hoses for connection.
•  Secure hoses and tubes against slippage and avoid kinks.
• Check hoses, tubes and bath at regular intervals for pos-

sible material fatigue (cracks/leaks).

• Mains cable should not get in contact with hot parts 

and fluids.

• If you are using plastic bath, observe the permitted 

working temperature range and fluids.

Do not start up the device if:

- It is damaged or leaking
- Cable (not only supply cable) is damaged.

• Be careful when filling a hot bath.

At high operating tempera-
tures, the temperature of hous-
ing parts, surfaces and tubes 
can exceed 70 ° C.

Safety instructions

NOTICE

WARNING

WARNING

DANGER

WARNING

Содержание CF.EG28.N10.80.8

Страница 1: ...20000004066b IKA ICC basic ICC basic_062017 ...

Страница 2: ...signation 1 Operator panel and display 2 Mains switch 3 Power socket 4 Buoyage 5 Heater 6 Stand 7 Pt 100 Pt 1000 temperature sensor 8 Handle 9 Fan 10 RS 232 port 11 USB port 12 Clamp 13 Cable clip 8 2 Fig 2 3 10 7 11 9 12 13 ...

Страница 3: ...and display 9 Commissioning 10 Setting the safety temperature 10 Menu settings 11 Menu structure 11 Basic guide to setting the menu options 11 Operating mode MODE 11 Maximum temperature HI T 11 Minimum temperature LO T 11 Maximum speed HI R 11 Minimum speed LO R 11 Fluid type FLUI 11 The type of temperature PID control arithmetic AUTO 11 Alarm and key tone BEEP 12 Calibration CALI 12 Interface and...

Страница 4: ... EN ISO 12100 Explication of warning symbols Indicates an extremly hazardous situation which if not avoided will result in death serious injury Indicates a hazardous situation which if not avoided can result in death serious injury Indicates a potentially hazardous situation which if not avoided can result in injury Indicates practices which if not avoided can result in equipment damage Indicates ...

Страница 5: ...nhibit fire or explosions or comprehensive monitoring equipment Process pathogenic material only in closed vessels under a suitable fume hood Please contact IKA application support if you have any question If the OFF switch is not within reach when device is operating an EMERGENCY STOP switch that can be easily accessed must be installed in the work area A laboratory circulator heats and circulates...

Страница 6: ... the heater The temperature of the heater can be very high After a power failure during operation the device may start automatically depending on operating mode Transport the device with care Do not transport or empty the bath while it is still hot This may result in accidents especially scalding injuries In order to prevent the power cable from falling into the fluid the mains cable must be se cur...

Страница 7: ...d by third parties Unpacking Delivery scope ICC baisc Mains cables Screwdriver use for safety circuit Operating instructions Warranty card Preparations Setting up Securely fix the immersion circulator on the bath Place the device on an even stable clean nonslip dry and fireproof surface Keep at least 20 cm of open space at the front and rear side When a plastic bath is used please ensure that the he...

Страница 8: ...bi distilled water are unsuitable for operation due to corrosive properties of these media High purity water and distillates are suitable as a medium after adding 0 1 g soda Na2CO3 sodium carbonate per liter of water o Note The temperature limit values are adjustable in accordance with the fluid used Nomenclature for IKA fluids CF EG28 N10 80 8 1 Classification HF Heating Fluid CF Cooling Fluid UF ...

Страница 9: ...avigation selecting the settings in the menu G OK Pump button Start stop the pump function Confirm the menu settings H Pump button Increase the pump speed setting value Change the value setting in the menu I Pump button Decrease the pump speed setting value Change the value setting in the menu J Menu button Press it once menu option is display Press it a second time back to the working screen K LED...

Страница 10: ...mode mode g Safety temperature safe i h After that following two screens appear alternatively Then the device enters standby status and is ready for op eration Change the temperature setting with Temp button E or Temp button F Change the pump speed setting with Pump button H or Pump button I In standby status activate the heating function by pressing the Temp button D The device start heating func...

Страница 11: ...ptions Enter the menu by pressing Menu button J Press the Temp button E or Temp button F to change the menu options Press the Pump button H or Pump but ton I to change the value settings in the menu Confirm the menu settings by pressing the OK Pump button G Operating mode MODE Mode A After power on power failure no automatic restart of functions Mode B After power on power failure automatic re star...

Страница 12: ...time Td s is the derivative time and de termines how strongly the rate of change of the control deviation affects the control variable Td compensates for rapid control deviations A high Td means a smaller and slower effect on the control variable Td values that are too small can lead to instability of the controller AUTO 0 For special requirements AUTO 0 can be used with manual adjustment of the c...

Страница 13: ...ss procedure data transfer from the stirrer machine to the computer takes place only at the computer s request Command syntax and format The following applies to the command set Commands are generally sent from the computer Master to the device Slave The device sends only at the computer s request Even fault indications cannot be sent spontaneously from the device to the computer automation system...

Страница 14: ...et the pump speed XXX OUT_SP_42 n Set the WD safety speed with echo of the set defined value OUT_WD1 n Start the watchdog mode 1 and set the time for the watchdog to n 20 1500 seconds Echo of the Watchdog time During a WD1 event the heating and pump functions are switched off This command needs to be send within the watchdog time OUT_WD2 n Start the watchdog mode 2 and set the watchdog time to n 20...

Страница 15: ...nly after it has been cleaned and is free from any materials which may constitute a health hazard For repair please request the Decontamination Cer tificate from IKA or download printout of it from the IKA website www ika com If you require servicing return the device in its original packaging Storage packaging is not sufficient Please also use suitable transport packaging Error codes Any malfuncti...

Страница 16: ...t the service department Send the device for repair including a short description of the fault Accessories Tubing hoses adapters and couplings LT 5 20 Metal hose isolated M16 x 1 LT 5 21 PTFE hose isolated M16 x 1 H PVC 8 PVC hose nominal width 8 H PVC 12 PVC hose nominal width 12 H SI 8 Silicone hose nominal width 8 H SI 12 Silicone hose nominal width 12 Tubing Insulations ISO 8 Insulation 8 mm I...

Страница 17: ...s III FL suitable for flammable and non flammable fluids Safety circuit adjustable C 0 160 Safety temperature display LED Heating capacity W 2000 230 VAC 1000 115 VAC 760 100 VAC Pump speed adjustable rpm 1000 3200 Max pump pressure suction bar 0 3 0 2 Max flow rate at 0 bar l min 18 Sub level protection Yes Interface USB RS 232 Permitted on time 100 Protection class according to EN 60529 IP 21 Protec...

Страница 18: ...be liable for freight costs The warranty does not cover worn out parts nor does it apply to faults resulting from improper use insufficient care or maintenance not carried out in accordance with the instructions in this operating manual Pump performance curve Fig 9 Pump performance curve measured with water Measurements done according DIN 12876 2 with water at 20 C pump in a closed loop circuit 0 ...

Отзывы: