16
Err 09
Pump off
Heating off
Device internal temperature is too
high
- Check the ambient temperature and let the de-
vice cool down
Err 10
Pump off
Heating off
PC communication failure
- Check communication cable
Err 11
Pump off
Heating off
Temperature difference between
control sensor and safety sensor is
too much
- Check safety temperature circuit and bath fluid
Err 12
Pump off
Heating off
Safety temperature alarm
- Check the bath temperature measurement
Err 13
Pump off
Heating off
Heater switched off by safety
circuit
- Check safety temperature set value, fluid level
Err 14
Pump off
Heating off
Fan error
- Check the fan and clean the grids at the rear side
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department;
- Send the device for repair, including a short description of the fault.
Accessories
•
Tubing, hoses, adapters and couplings:
LT 5.20
Metal hose (isolated M16 x 1)
LT 5.21
PTFE hose (isolated M16 x 1)
H.PVC.8
PVC hose (nominal width 8)
H.PVC.12
PVC hose (nominal width 12)
H.SI.8
Silicone hose (nominal width 8)
H.SI.12
Silicone hose (nominal width 12)
•
Tubing Insulations:
ISO. 8
Insulation (8 mm)
ISO.12
Insulation (12 mm)
•
Bath vessels:
IB eco 8
Plastic bath (8 litres)
IB eco 18
Plastic bath (18 litres)
IB pro 9
Stainless steel bath (9 litres)
IB pro 12
Stainless steel bath (12 litres)
IB pro 20
Stainless steel bath (20 litres)
•
Bridges and Covers:
BS.ICC
Small bridge (for
IB eco 8
,
IB pro 9
)
BL.ICC
Large bridge (for
IB eco 18
,
IB pro 12, IB pro 20
)
CS.ICC
Samll cover (for
IB eco 8
,
IB pro 9
)
CM.ICC
Medium cover (for
IB pro 12
)
CL.ICC
Large cover (for
IB eco 18
,
IB pro 20
)
•
Additional accessories:
PCS.ICC
Pump set
PC 1.1
Cable (RS 232)
USB 2.0 cabel A - Micro B
Labworld
soft
®
See more accessories on
www.ika.com
.