17
Technical data
Operating voltage
VAC
230 ± 10 %
115 ±10 %
100 ± 10 %
Frequency
Hz
50/60
Max. input power
W
2100 (230 VAC)
1100 (115 VAC)
860 (100 VAC)
Working temperature range (RT+10 °C at 1000 rpm)
°C
RT + 10 ... 150
Operating temperature range (with forced cooling)
°C
- 20 ... 150
Temperature stability – Internal temperature
control 70 °C, water (according to DIN12876)
K
± 0.02
Temperature control
PID
Temperature measurement, absolute accuracy
K
± 0.2
Temperature setting
Button
Temperature setting resolution
K
0.1
Temperature display
LED
Temperature display resolution
K
0.1
Classification according to DIN12876-1
Class III (FL) suitable for flammable and non-flammable fluids
Safety circuit (adjustable)
°C
0 ... 160
Safety temperature display
LED
Heating capacity
W
2000 (230 VAC)
1000 (115 VAC)
760 (100 VAC)
Pump speed (adjustable)
rpm
1000 ... 3200
Max. pump pressure / suction
bar
0.3 / 0.2
Max. flow rate (at 0 bar)
l/min
18
Sub-level protection
Yes
Interface
USB, RS 232
Permitted on-time
%
100
Protection class according to EN 60529
IP 21
Protection class
I
Excess voltage category
II
Contamination level
2
Permitted ambient temperature
°C
+ 5 ... + 40
Permitted ambient humidity
%
80
Dimension (W x D x H)
mm
145 x 200 x 340
Weight
kg
3.75
Operation at a terrestrial altitude
m
max. 2000
Note
:
Complies to EN61000-3-11 subject to conditional connection: Zmax = 0.282 Ω. If necessary, consult your electricity supplier.