20
10.5.7 Low flow cut-off
►
Select [LFC] and set the limit�
10.6 Service functions
10.6.1 Read min/max values
Read minimum or maximum measured values:
►
Select HI or LO
HI = maximum volumetric flow, LO = minimum volumetric flow
Delete memory:
►
Select HI or LO�
►
Press and hold [Set] until [----] is displayed�
►
Briefly press [Mode/Enter]�
It makes sense to delete the memories as soon as the unit operates under normal
operating conditions for the first time�
10.6.2 Restoring the factory settings
►
Select [rES]�
►
Press and hold [Set] until [----] is displayed�
►
Briefly press [Mode/Enter]�
11 Operation
Correct operation and compliance with the measurement accuracy can only be en-
sured if the environmental conditions specified in the "Technical data" are adhered
to� Ensure that the maximum pressure range, measuring range and permitted
operating temperature are not exceeded�
After power on and expiry of the power-on delay time (approx� 0�5 s) the unit is in
the Run mode (= normal operating mode)� It carries out its measurement and eval-
uation functions and generates output signals according to the set parameters�
→ 8 Operating and display elements
• During the power-on delay time the outputs are switched as programmed: ON
for NO function (Hno / Fno), OFF for NC function (Hnc / Fnc)�
• If output 2 is configured as analogue output, the output signal is at the final
value of the measuring range during the power-on delay time�
11.1 Reading of the set parameters
►
Press [Mode/Enter] until the requested parameter is displayed�