8
6 Installation
►
The rules and regulations for the installation and operation of com-
pressed air equipment must be observed�
►
Ensure that no pressure is applied to the installation during mounting�
6.1 Installation location
• Behind the cold dryer / near the load�
• If compressed air is fed into the main pipe through parallel pipes, the unit
should be mounted in the main pipe�
• Installation after the maintenance unit is also possible (if oil is used for the
loads, the units must be mounted before the oiler)�
• Ensure sufficient clear space (min� 120 mm) for installation and removal of the
sensor�
6.2 Installation conditions
To meet the specified measurement accuracy, the defined inlet and outlet pipe
lengths must be adhered to:
• Inlet pipe length: 15 x pipe diameter�
• Outlet pipe length: 5 x pipe diameter�
In case of disturbances on the inflow side additional calming sections (B) are
recommended�
changes of the
pipe diameter
B = 5 x pipe diameter (D)
90° elbow
B = 5 x pipe diameter (D)
two 90° elbows, one plane
B = 10 x pipe diameter (D)
two 90° elbows, two planes
B = 15 x pipe diameter (D)
valve, slide
B = 35 x pipe diameter (D)