6.3 Installation position
• Permitted installation positions: pipe length vertical, any position (fig� 1, 2), pipe
length horizontal, unit vertical (fig� 3, 4), unit on side, pipe length left (fig� 5)�
• Avoid the installation position in fig� 6 (unit on side, pipe length right)� If the flow
rate is low, the specified measurement accuracy cannot be adhered to�
F
F
1
2
3
5
4
6
F = flow direction
6.4 Install the pipe section into the pipe
IN
OUT
►
Loosen the screws (1) on the flanges, remove the flanges (2) from the measur-
ing section (4)�
►
Remove the seals (3) from the flange slots�
►
Weld the flanges into the pipe�