Min� diameter of the bypass and still pipe: d = 100 mm�
The upper access to the steady area (fig� 6-5: A / B) has to be above the
max� level� The lower access (fig� 6-5: C / D) or the area with perforated
sheet has to be below the min� level� This ensures that neither foam nor
turbulence impact the sensor zone� When perforated sheets or the like are
used, soiling (e�g� by metal swarf, particles, ���) can also be avoided�
Fig. 6-5
d
d
A
B
C
D
6.2 Installation of the probe
The rod is not included in the scope of supply� It has to be ordered separately
(→ 3 Items supplied)
�
Fixing of the rod:
►
Screw the rod to the unit and tighten it�
Recommended tightening torque:
4 Nm�
For ease of installation and removal the rod
connection can be rotated without restriction�
Even if rotated several times there is no risk
of damage to the unit�