TITAN HD USER MANUAL
92
9-3-1 Emissivity
The correct emissivity value is important in order to receive the most accurate
temperature measurement. Emissivity of a surface can have a large effect on
temperatures the imager observes.
To more accurately measure materials with a low emissivity, an emissivity
correction is necessary. Adjusting the emissivity setting allows the imager to
calculate a more accurate estimate of the actual temperature.
Emissivity is set directly in the imager. The emissivity displays on the LCD screen
as ϵ=x.x
See
Section 8-17-1-3 ROI Regional Environmental Parameters
and
Section 8-17-
2 Global Environmental Parameters
to learn how to adjust the emissivity value
of an ROI.
All objects radiate infrared energy. The amount of energy radiated varies based
on the actual surface temperature and the surface emissivity of the object. The
Thermal imager senses the infrared energy from the surface of the object and
uses this data to calculate an estimated temperature value.
Many common objects and materials such as painted metal, wood, water, skin,
and cloth are particularly good at radiating energy and it is easy to get relatively
accurate measurements. For surfaces that are good at radiating energy (high
emissivity), the emissivity factor is >=0.90. This simplification does not work on
shiny surfaces or unpainted metals as they have an emissivity of <0.6. These
materials are not good at radiating energy and are classified as low emissivity.
A CHART CONTAINING COMMON EMISSIVITY VALUES APPEARS ON THE
NEXT PAGE.
SURFACES WITH AN EMISSIVITY OF < 0.60 MAKE RELIABLE AND CONSISTENT
MEASUREMENTS PROBLEMATIC. THE LOWER THE EMISSIVITY, THE GREATER
POTENTIAL FOR ERROR WITHIN THE DEVICE’S CALCULATIONS. ERRORS MAY
STILL OCCUR EVEN WHEN ADJUSTMENTS ARE PERFORMED PROPERLY.
The device features a built-in emissivity value table. Select a value as follows:
1.
Using the touchscreen, tap the settings icon to open the Settings Menu.
2. Tap the Environmental Parameters icon:
3. Tap the drop-down arrow next to Emissivity of the ROI to be adjusted. A
table appears.
4. Press-and-hold on the touchscreen; then, drag to scroll through the options.
5. Tap an option to select it.
6. Press the Previous button continuously to exit the menu.
A chart containing common emissivity values appears on the next page.
Содержание TITAN HD
Страница 2: ...TITAN HD USER MANUAL 2 Revision History 01 2023 001 Document created...
Страница 11: ...TITAN HD USER MANUAL 11 2 9 Training To read about infrared training visit https infraredtraininginstitute com...
Страница 98: ...TITAN HD USER MANUAL 98 11 6 Camera out of focus Adjust focus ring 11 7 Memory card error Insert memory card...