SL193-011342 ED00 10-2018
47
Software updates
For a CPU software upgrade, it is necessary to use the special serial cable
supplied by IceTeam1927 (IC577400056), connect it to the CPU through
the Serial port (see electronic board layout) and to the USB port or RS232
of any Personal Computer (PC). The software Upgrade can be done by
transferring the new software from the computer to the CPU using M16C Flasher. Once you
have connected your PC (RS232 or USB) to the CPU (serial), switch on the machine. Once
M16C Flasher has been installed:
-
Start the program and press the Settings button.
-
Configure as follows:
This configuration is necessary only the first time the program is used.
-
Close the Settings dialog.
-
Press the arrow next to the Settings
button and select Prog new file…
-
Select the P1-*.mot program.
After a software update there is no need to adjust the sensor offsets or the Trimmer Hot.
Содержание SL193-011342
Страница 1: ...Electronic Manual English Version Ed00 10 2018 SL193 011342 P1...
Страница 2: ......
Страница 4: ...File name SW version Date Changes to previous version P1 1_00 mot 1_00 22 10 2018 First version...
Страница 6: ...XIX C Ohm table 48 XX Sensor calibration after electronic board replacement 50...
Страница 12: ...SL193 011342 ED00 10 2018 9 Figure 3 PR Left Side PR EVRC TEC EVRC TEV EVRC Figure 4 5 TEC TEV...
Страница 38: ...SL193 011342 ED00 10 2018 35 Step Contents U14 Wash hour Setting of display hour of the Wash today alarm...
Страница 52: ...SL193 011342 ED00 10 2018 49 85 1508 90 1557 9 95 1608 7 100 1660 4 105 1713 110 1766 5 115 1820 9...
Страница 57: ...REALIZED by VERIFIED by APPROVED by Francesco Curatolo Francesco Curatolo Salvador Carbonell Salvador Carbonell...
Страница 58: ......