![IceCure ProSense FAS3000000 Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/icecure/prosense-fas3000000/prosense-fas3000000_user-manual_610296054.webp)
IceCure™ Medical Ltd.
DMS-7064 rev. C
ProSense™
Confidential
User Manual
54
European User Manual
Electrical requirements
For countries in which local line voltage is 100-127 VAC
The
ProSense™
cryoablation system is pre-wired for the local line voltage as specified by
the user. Accordingly, the unit will require a line supply of 100-127 VAC, 50/60 Hz, 12 A
single phase.
The unit is grounded through the power cable that is plugged into the wall power outlet.
Good grounding is essential for safe operation. The main circuit breaker for the unit is
0.1 A. The circuit breaker is located in the main chassis.
For countries in which local line voltage is 220-240 VAC
The
ProSense™
cryoablation system is pre-wired for the local line voltage as specified by
the user. Accordingly, the unit will require a line supply of 220-240 VAC, 50/60 Hz, 7A
single phase.
The unit is grounded through the power cable that is plugged into the wall power outlet.
Good grounding is essential for safe operation. The main circuit breaker for the unit is 0.1
A. The circuit breaker is located in the main chassis.
Shipment components
The
ProSense™
cryoablation system is packed and shipped in separate units as follows:
•
Main system
•
User manual
•
Two empty Dewars
•
Accessories: foot pedal (Not available in some regions, e.g. China)
The following products are not supplied with the system and should be ordered
independently:
•
Sterile single use Cryoprobes
•
Sterile single use Introducers
•
Sterile single use Temperature sensors
•
Sterile sleeves for the hose
•
Sterile cover for the cryoprobes cradle
•
Holder (Optional - Not available in some regions, e.g. China)
•
Liquid nitrogen
•
Disinfecting wipes
•
Dewar for liquid Nitrogen (Large volume)
•
Liquid Nitrogen withdrawal device