European User Manual
Danger - explosion and fire hazard
Warning
The
ProSense™
cryoablation system includes electronic devices that may emit
sparks, and therefore, should not be operated in the presence of flammable
materials.
ProSense™
cryoablation system should be kept away from flammable fumes, e.g.,
flammable anesthetics or volatile substances. An easily accessible fire extinguisher in the
vicinity of the unit is recommended.
Opening the Dewar compartment
Caution
Never open the Dewar compartment while running a cryoablation
procedure. Do not use the system while the Dewar compartment is open.
Warning
To avoid spillage and cold burns, do not transfer a Dewar containing liquid
nitrogen unless it is covered with its designated lid.
Warning
Follow the specific instructions on how to fill and transfer a Dewar as
described in section 5.2.6.1.
Liquid nitrogen Safety
Warning
Liquid nitrogen may cause, if there is not sufficient ventilation in the room,
serious injury or burn if handled improperly. Local laws and safety rules
regarding the maintenance and handling of liquid nitrogen Dewars should
always be observed. Maintenance of liquid nitrogen Dewars should be
performed by authorized personnel only.
Warning
To avoid spillage and cold burns, do not transfer a Dewar containing liquid
nitrogen unless it is covered with its designated lid.
Warning
If there is not sufficient ventilation in the room, the
ProSense™
cryoablation
system cannot be used due to risk of suffocation due to increased levels of
nitrogen in the room.