European User Manual
Warning
Do not start the
Freeze
step before the cryoprobe active area is within the
target tissue.
Voltage and power ratings
Caution
Verify that the
ProSense™
cryoablation system complies with the local
voltage line.
Installation and Setup
Caution
The
ProSense™
cryoablation system must be unpacked, installed, and tested
by an
IceCure™
Medical technician or by a technician authorized by
IceCure™
Medical to perform installation and testing on
ProSense™
system.
Caution
After positioning the main chassis, lock the front roller brakes. Failure to do
so may result in damage to the system or to other equipment in the clinic
room.
Warning
The cryoablation system should be operated in an adequately ventilated
room. Failure to do so may result in risk of suffocation due to increased levels
of nitrogen in the room.
Caution
There are no user-serviceable parts in the system. Refer all service issues to
IceCure™
Medical’s Customer Service Department.
Caution
Make sure the time is set according to local time zone before executing a
cryotherapy procedure.
Proper use
Warning
Do not disconnect the cryoprobe from the system during
Freeze
,
Extraction
,
or
Thaw
unless you are instructed to do so by the system.