IceCure™ Medical Ltd.
DMS-7064 rev. C
ProSense™
Confidential
User Manual
105
European User Manual
Removing the cryoprobe from the cryohandle
When a Hose-holder is used during the procedure, safely disengage the hose from the
Holder device.
After removing the cryoprobe from the target tissue, and the message that it is safe to
disengage the cryoprobe is displayed, detach the cryoprobe from the cryohandle as
follows:
1.
Unscrew the used cryoprobe from the cryohandle and dispose of it appropriately.
2.
Remove the single-use sterile cover from the cryohandle.
3.
Close the cryohandle with the covering plug.
Figure 109: Safe to Remove cryoprobe
Warning
Dispose of the cryoprobe, temperature sensor and cryohandle sleeve in
accordance with institutional policy.
Do not reuse, do not re-sterilize.
Disassembling the support device - Hose-Holder
If a Hose-Holder accessory is used as a support device, safely disengage the hose from the
Holder arm.
Disassembling the temperature sensor
If a temperature sensor was used, pull the gray part of the cable connector backwards to
release it, and then pull the cable connector away from the front panel, dispose of it
appropriately.