The temperature sensor is not indicated for measuring the body temperature, but only
for getting temperature indication of some tissue during cryoablation.
Warning
Do not rely solely on the temperature sensor measurement. Always monitor
the procedure using Ultrasound or other appropriate imaging systems.
The temperature sensor and its components are depicted in the following figure.
Figure 36: Temperature sensor and its components
If you decide to use a temperature sensor, connect it per the following instructions while
maintaining sterility:
1.
Remove the temperature sensor from the sterile package.
2.
Connect the cable connector to the panel connector on the upper right corner of
the main chassis while keeping the connector key facing the groove.
3.
The measured temperature from the sensor tip will appear on the right lower
corner of the screen.
4.
Under imaging guidance, insert the tip of the temperature sensor into the tissue
you want to measure.
5.
At the end of the procedure, gently remove the temperature sensor from the
tissue and discard it.