IceCure™ Medical Ltd.
DMS-7064 rev. C
ProSense™
Confidential
User Manual
120
European User Manual
Problem
Probable Cause
Action
Cryoprobe mechanical
damage
Bent or pinched cryoprobe.
1. Press the red emergency stop
button
2. Wait for 60 seconds.
3. If the cryoprobe is in the
patient’s body, wait for passive
thaw if necessary and remove
the cryoprobe gently.
•
Make sure that the cryoprobe is
removed intact.
4.
Make sure that the tip is
pointed in a safe direction and
then securely disengage the
cryoprobe.
5.
The system should not be
turned on until disengaging the
cryoprobe from the cryohandle.
The cryoprobe should be
screwed out of the cryohandle.
6.
Close the handle plug on the
cryohandle, by screwing it in.
7.
Start the system, using the on-
off switch on the back side of
the system.
8.
If required, wait for the Dewar
to be lowered down and
perform a new procedure with
full Dewar and a new cryoprobe
while maintaining sterility.
Freeze/Extraction
icon
does not initiate the
operation
Faulty icon or operator misuse.
Try pressing the
Freeze/Extraction
icon again. It is only activated if
pressed for 1 second.
Check the user manual to ensure
that you can indeed use the icon in
the present screen.
If the problem persists contact
IceCure™
Medical service.
LED does not light up
Burned out LED, or it should not
be lighting up at given time.
Check user manual to verity that the
LED should light up.
If yes, contact
IceCure™
Medical
service.
Ice formation on the
hose, the cryohandle
and sterile sleeve
Low temperature and long
freezing steps may cause moisture
to form ice particles.
This is not an operational problem.
Wipe OFF accumulated ice with
clean sterile gauze.
Liquid nitrogen leaks
from cryoprobe,
cryohandle or system
during the cryotherapy
procedure.
Cryoprobe is not well connected
or pipes are faulty.
Press the
Extraction
button or
emergency button and contact
IceCure™
Medical service.