22
Touch Screen Calibration
The touch screen will require occasional recalibrating. This is a very quick and simple procedure.
If you are noticing difficulty in the touch screen responding when touched,
do not
press harder or use a
hard or pointed object to get the screen to respond.
Display
Example
#12
Display
Example
#13
Display Example #14
Display Example #15
Door Lift Calibration
The Door Lift Calibration is used to determine the travel time for the lift motor to raise and lower the
platform. The calibration values are used to provide more accurate preheat and cooling times. Touch
the “
Calibrate
” button and the lift will raise and lower. The travel times will be displayed and
recorded in memory. Touch the “
Back
” button to return to the calibration screen.
Display Example #16
CALIBRATION
3 touches left
IMPORTANT.
Now touch screen
calibration will be
performed. Touch points
EXACTLY at the positions
shown by arrows.
CALIBRATION
3 touches left
CALIBRATION
3 touches left
Door Calibration
Door Close Time:
13.0 s
Door Open Time:
13.5 s
Calibrate
Back
Содержание SUMMIT